HB Ad Slot
HB Mobile Ad Slot
PMRA Issues Interim Order to Regulate Certain UV Radiation-Emitting and Ozone-Generating Devices Under the Pest Control Products Act
Monday, June 14, 2021

On June 7, 2021, Health Canada’s Pest Management Regulatory Agency (PMRA) announced the issuance of its Interim Order Respecting Ultraviolet Radiation-emitting Devices and Ozone-generating Devices under the Pest Control Products Act (Interim Order), setting forth new requirements for certain devices claiming to control, destroy, make inactive, or reduce the level of bacteria, viruses, and other micro-organisms that are human pathogens. PMRA also issued an “Explanatory Note” and a Questions and Answers document regarding the Interim Order.

PMRA states that it issued the Interim Order following the increased sale of ultraviolet (UV) radiation-emitting and ozone-generating devices such as lights and wands in Canada since the COVID-19 pandemic. These devices are marketed to kill bacteria and viruses, including SARS-CoV-2, the virus that causes COVID-19. According to PMRA, it has not received enough evidence to confirm that UV radiation-emitting and ozone-generating devices are safe for users and the public, or that they are effective.

Thus, PMRA is now requiring that companies register certain UV radiation-emitting devices and ozone-generating devices before they may be sold or used in Canada. In its Explanatory Note, PMRA states:

By bringing certain UV and ozone-generating devices under the [Pest Control Products Act (PCPA)], they need to be registered or otherwise authorized in order to be on the Canadian market. An application to register a pest control product must be submitted to Health Canada in the form and manner directed by the Minister and must include any information and other material that is required by the Pest Control Products Regulations to accompany the application. Applications to register devices consist of a number of information and data requirements, including a cover letter stating the purpose of the application, an application and fee estimate forms, the proposed English and French product labels, as well as data to support the safety and efficacy of the device. A registration will be granted under the PCPA if the Minister considers that the health and environmental risks and the value of the device are acceptable after any required assessments.

The Interim Order clarifies that certain UV radiation-emitting devices and ozone-generating devices claiming to kill bacteria and viruses are not subject to the regulatory requirements of the PCPA and its Regulations. These include:

  • Devices that are manufactured, represented, distributed, or used to control, destroy, or inactivate viruses, bacteria, or other micro-organisms that are human pathogens for use in swimming pools, spas, or wastewater treatment systems;

  • Devices that meet the definition of “device” in Section 2 of the Food and Drugs Act and are classified as a Class II, III, or IV medical device under the Medical Devices Regulations; and

  • UV radiation-emitting devices that satisfy the following conditions:

    • The device is certified by a standards development organization accredited by the Standards Council of Canada as meeting the applicable Canadian electrical safety requirements;

    • The certification mark of the standards development organization appears on the label of the device;

    • Any efficacy claim that is made in respect of the device is only a claim of supplemental sanitization;

    • No express or implied reference to prevention, treatment, or mitigation of disease is made in respect of the device;

    • The device has at least one of the following mechanisms:

      • A mechanism that locks the device during operation, or

      • A mechanism that automatically shuts off the device if it is opened during operation; and

    • The UV lamp is fully shielded or enclosed in the device in a manner that prevents access to it by users of the device and prevents exposure to UV radiation.

For UV radiation-emitting devices that satisfy these conditions, there are additional labeling requirements for the display panels and operating manual.

Discussion

This Interim Order changes significantly the requirements applicable to these types of devices that did not previously require registration. PMRA is providing a 30-day transition period after the Interim Order is issued before companies must comply with these requirements. PMRA also is advising Canadians to “stop using UV lights and wands that claim to disinfect against the virus that causes COVID-19 especially if the product is for use on the skin.”

Companies that produce devices that may be subject to this Interim Order should review the new requirements and exemptions carefully. PMRA states that unregistered or unauthorized devices are prohibited and may be subject to compliance and enforcement action.