FDA Issues Final Rule to Regulate Laboratory-Developed Tests as Medical Devices


On May 6, 2024, the U.S. Food and Drug Administration published its anticipated final rule concerning oversight of laboratory-developed tests (LDTs). The final rule (i) affirms the FDA’s position that in vitro diagnostic products (IVDs) are devices under the Federal Food, Drug, and Cosmetic Act (FDCA), including when the manufacturer of the IVD is a laboratory; (ii) sets out FDA’s enforcement discretion phaseout policy for LDTs in stages; (iii) applies “targeted” enforcement discretion for certain categories of LDTs; and (iv) reiterates that certain other categories of LDTs, which receive either full enforcement discretion or no enforcement discretion at all, are not subject to the phaseout policy. Concurrently with the publication of the final rule, FDA also issued two draft guidance documents: (i) Consideration of Enforcement Policies for Tests During a Section 564 Declared Emergency, and (ii) Enforcement Policy for Certain In Vitro Diagnostic Devices for Immediate Public Health Response in the Absence of a Declaration under Section 564.

FDA defines LDTs as a subset of IVDs designed, manufactured, and used within a single laboratory certified under the Clinical Laboratory Improvement Amendments (CLIA) to perform high-complexity testing. FDA has stated that increased oversight of LDTs is necessary to protect patients and ensure the accuracy of diagnostic tests. According to FDA, many modern LDTs rely on increasingly high-tech instrumentation or software and are frequently used to guide critical decisions, yet scientific data collected over the years indicates that IVDs offered as LDTs do not always produce accurate or reliable results. Therefore, FDA believes that LDTs should be required to meet FDA requirements for medical devices to ensure patient safety.

Enforcement Discretion Phase-Out Policy for LDTs

Under the final rule, FDA will gradually phase out its exercise of enforcement discretion for LDTs in five stages over the next four years. Each stage contains a deadline for achieving certain regulatory requirements, such that by the Stage 5 deadline, all LDTs subject to oversight must be in full compliance with medical device requirements under the FDCA. All LDTs must comply with Medical Device Reporting requirements, reports of correction and removals, registration and listing requirements, and labeling requirements as part of the phaseout policy unless expressly excluded.


*As long as a premarket submission has been received by this date, FDA states it will continue to exercise enforcement discretion for the LDT for the pendency of its review.

Targeted Enforcement Discretion

Notably, the final rule applies “targeted” enforcement discretion to the following categories of LDTs as part of the phaseout policy. According to FDA, these limited exceptions were added to the final LDT rule in response to public comments on the proposed LDT rule.

LDTs Excluded from Phaseout Policy

The FDA intends to exercise full enforcement discretion for certain categories of LDTs, which continue to be exempt from all premarket review and quality requirements for medical devices under the FDCA or facility registration, as long as the LDTs are designed, manufactured, and used within a CLIA-certified, high-complexity laboratory:

LDTs Not Subject to General Enforcement Discretion

FDA also states that certain other categories of LDTs are required to comply with device requirements under the FDCA:

GT Takeaways from Final LDT Rule

While the phaseout policy and the new targeted enforcement discretion categories provide an extended “on-ramp” for affected LDT manufacturers, as FDA underscores in the final rule, any enforcement discretion policy is not law and can be changed by FDA at any time, without additional rulemaking, as long as the changes are “consistent with good guidance practices.” Namely, if FDA is concerned with the safety and effectiveness of any LDT already on the market, it can still require the performing laboratory to respond to inquiries and/or obtain 510(k) clearance for the LDT. FDA asserts that it will explicitly request submission of the labeling for currently marketed LDTs, to use with other information submitted under the phaseout policy, to identify LDTs that “specifically raise concerns.” This scrutiny may also be triggered by publicly available information, or a complaint, followed by an investigation conducted by FDA. Stakeholders should factor this uncertainty into their compliance planning efforts under the phaseout policy.

The final rule also leaves several items up for interpretation, especially with regard to which types of modifications would trigger the need for premarket review and how the labeling submission requirements apply to LDTs over which FDA intends to exercise targeted enforcement discretion.

Finally, while submitting an LDT for premarket review prior to the start of Stages 4 or 5 of the phaseout policy (as applicable) has the advantage of extending enforcement discretion pending a decision from FDA, the earlier submitters will need to navigate premarket review pathways (e.g., 510(k), de novo), which may involve uncertainty as to the availability of predicate devices, due to the differences between IVDs generally versus IVDs manufactured by laboratories.

Considerations for Laboratories

Laboratories must ensure that they understand all distinctions and nuances under the final rule and comply with the applicable regulations under this phaseout policy within the prescribed timeframes. The two most immediate compliance requirements for LDT manufacturers, which must be fully implemented by May 6, 2025, relate to the filing of medical device reports (21 C.F.R. Part 803), as well as notices of corrections and removals (21 C.F.R. Part 806). Prior to this first deadline, laboratories may wish to amend all policies and procedures related to the development and use of LDTs to ensure (i) proper documentation of all investigations, audits, reports, and corrective actions taken in response to any LDT-related incidents; and (ii) regular review and appropriate reporting of data related to adverse events, corrections, and removals.

Although after May 6, 2025, the next deadline is not until Feb. 2, 2026, the development and implementation of quality systems for complaint files is a multi-step process involving complex documentation and registration requirements necessitating robust organizational planning to implement. Therefore, while medical device reporting and correction and removal notices should be the first priority, laboratories should begin preparing for QMSR compliance no later than early 2025.


©2025 Greenberg Traurig, LLP. All rights reserved.
National Law Review, Volume XIV, Number 131